נחל שניר 8, יבנה

PyroCube M

Measurement Specifications

Temperature Range
100°C to 600°C

Provided emissivity > 0.3, measurements below 100°C are possible with reduced stability.

Analogue Output Scale
Factory set to:
4 mA = 50°C
20 mA = 600°C
Adjustable via optional touch screen module or RS232
Response Time (95%)
1 ms (adjustable via averaging function)
Measured Spot Diameter
From 1.0 mm depending on measurement distance and choice of optics (see Field of View Diagrams below)
Target Sighting
Red LED built-in as standard on all models, shows the position and size of the measurement area. Switchable on/off.
Accuracy of Measurement
± 3°C or 1%, whichever is greater
Repeatability
± (0.2% + 2°C)
Temperature Resolution
0.5°C
Emissivity Setting
Adjustable via RS232C or optional touch screen interface
0.05 to 1.0